BRAND : MEDI WAVES
MODEL : MV
Device Specifications
Parameter Specification
Rated Voltage AC120V or AC230V
Rated Frequency 50Hz / 60Hz
Power Consumption 1200VA ±10%
Fuse T10AH when AC120V or T6.3AL when AC230V
Protection Class Class 1, Type CF
Leakage Current In accordance with IEC601, Part2-2
Carrier Frequency 400㎑, 500㎑
Repeat Frequency 33㎑
Size (W × D × H) 410mm × 388mm × 160mm
Weight 12.5Kg
Operation Temperature 10℃ to 40℃
Storage Temperature -10℃ to 60℃
Humidity 20% to 95% RH
Operation Altitude 700mbar ~ 1060mbar
Operation Cycle 10 sec ON / 30 sec Idle
Cooling 1 inner fan
Output Power
Mode Output Power Load Carrier Frequency Crest Factor Duty Rate Repeat Frequency
Pure Cut 400W 200Ω 400kHz 1.6 100% Continuous
Blend 1 250W 200Ω 400kHz 2.4 70% 33kHz
Blend 2 200W 200Ω 400kHz 2.3 55% 33kHz
Blend 3 150W 200Ω 400kHz 2.8 45% 33kHz
Contact Coagulation 120W 200Ω 400kHz 3.3 40% 33kHz
Spray Coagulation 100W 300Ω 400kHz 5.9 8.0% 33kHz
Argon Coagulation 100W 300Ω 400kHz 5.9 8.0% 33kHz
Bipolar Standard 100W 100Ω 500kHz 1.3 100% Continuous
Bipolar Auto Start 100W 100Ω 500kHz 1.3 100% Continuous
Bipolar Forced 80W 100Ω 500kHz 2.7 30% 33kHz
Bipolar Cut 120W 200Ω 500kHz 1.6 100% Continuous
Bipolar Blend 100W 200Ω 500kHz 2.2 65% 33kHz
Output Power (Optional)
Mode Output Power Load Carrier Frequency Crest Factor Duty Rate Repeat Frequency
Soft Coagulation 80W 100Ω 400kHz 1.4 100% Continuous
Papillotomy 150W 200Ω 400kHz – – –
Polypectomy 150W 200Ω 400kHz – – –